Анализ прочностных характеристик цементно-винтовой фиксации мульти-юнит абатментов к дентальным имплантатам с плоскостным и коническим типом соединения в эксперименте

Analysis of strength characteristics of cement-screw fixation of multi-unit abutments to dental implants with flat and conical type of connection in experiment

Thumbnail
Открыть
118_124.pdf (315.2Kb)
Дата
2025
Автор
Прялкин, С. В.
Рубникович, С. П.
Бусько, В. Н.
Metadata
Показать полную информацию

Аннотации

Цель – провести анализ прочностных характеристик цементно-винтовой фиксации мульти-юнит абатментов к дентальным имплантатам с плоскостным и коническим типом соединения в эксперименте в условиях циклических нагружений, имитирующих функциональные нагрузки в ротовой полости. Методы. Оценивали прочностные характеристики цементно-винтового соединения (ЦВС) мульти-юнит абатментов к дентальным имплантатам из сплава Ti-6Al-4V диаметром 3,75 мм и длиной 11,5 мм с коническим и плоскостным типом соединения. Всего в эксперимент было включено 48 дентальных имплантатов - 24 с плоскостным и 24 с коническим типом соединения. В свою очередь каждые 24 имплантата были распределены на шесть групп в зависимости от экспериментальных условий: три группы при угле нагружения 22,5° и силе воздействия 100 Н с количеством циклов нагружения 100 000, 150 000 и 200 000; три группы при угле нагружения 45° и силе воздействия 100 Н с количеством циклов нагружения 100 000, 150 000 и 200 000. Результаты. При сравнительном анализе прочности экспериментальных образцов в условиях максимальной нагрузки (угол нагружения 45° и 200 000 циклов нагружения) во 2-ом положении импланта, как наиболее слабом элементе всей конструкции, выявлено, что при коническом типе глубина ЦВС снижается на 19 %, а степень зажатия винтов – на 73 % от изначального уровня до нагрузки, в то время как при плоскостном типе аналогичные параметры снижаются на 33 % и 80 % соответственно. Заключение. Установлено, что конический тип соединения является более надёжным нежели плоскостной тип, его использование позволяет повысить сопротивление нагрузкам, так как накопление усталости ортопедических конструкций при коническом мульти-юнит соединении возрастает в 1,2 раза, при плоскостном мульти-юнит соединении – в 1,5 раза. соединения, имитирующего проекцию 2-го премоляра) на 24 % по сравнению с плоскостным типом соединения
 
Purpose is to analyze the strength characteristics of the cement-screw fixation of multi-unit abutments to dental implants with a planar and conical type of connection in an experiment under conditions of cyclic loads simulating functional loads in the oral cavity. Methods. The strength characteristics of the cement-screw connection (CSC) of multi-unit abutments to dental implants made of Ti-6Al-4V alloy with a diameter of 3.75 mm and a length of 11.5 mm with a conical and planar type of connection were evaluated. A total of 48 dental implants were included in the experiment – 24 with a planar and 24 with a conical type of connection. In turn, every 24 implants were divided into six groups depending on the experimental conditions: three groups with a loading angle of 22.5° and a force of 100 N with a number of loading cycles of 100 000, 150 000 and 200 000; three groups at loading angle of 45° and impact force of 100 N with number of loading cycles of 100 000, 150 000 and 200 000. Results. In comparative analysis of the strength of experimental samples under maximum load conditions (loading angle of 45° and 200,000 loading cycles) in the 2nd position of the implant, as the weakest element of the entire structure, it was revealed that with the conical type, the depth of the CSC decreases by 19 %, and the degree of screw clamping – by 73 % from the initial level to the load, while with the planar type, similar parameters decrease by 33 % and 80 %, respectively. Conclusion. It has been established that the conical type of connection is more reliable than the planar type, its use allows you to increase the resistance to loads, since the accumulation of fatigue of orthopedic structures with a conical multi-unit connection increases in 1.2 times, with a planar multi-unit connection – in 1.5 times. compound simulating the projection of the 2nd premolar) by 24 % compared to the planar type of compound.
 

Библиографическое описание

Прялкин, С. В. Анализ прочностных характеристик цементно-винтовой фиксации мульти-юнит абатментов к дентальным имплантатам с плоскостным и коническим типом соединения в эксперименте / С. В. Прялкин, С. П. Рубникович, В. Н. Бусько // Медицинский журнал. – 2025. – № 4 (94). – С. 118–124.
URI
https://rep.bsmu.by/handle/BSMU/55941
Collections
  • Медицинский журнал. 2025. № 4(94) [28]
  • Научные публикации ученых БГМУ. 2025 [2986]

Белорусский государственный медицинский университет | Библиотека БГМУ | Контакты
 

 

ISSN 2521-6562 online

Просмотр

Весь РепозиторийРазделы и коллекцииДата публикацииАвторыНазванияТематикаЭта коллекцияДата публикацииАвторыНазванияТематика

Моя учетная запись

ВойтиРегистрация

Индексация

Google ScholarOpenDOARROAD ISSNrepositories.webometrics.info

Белорусский государственный медицинский университет | Библиотека БГМУ | Контакты